=== pH electrode and reference electrode design === Details of the fabrication and resulting microstructure of the glass membrane of the pH electrode are maintained as trade secrets by the manufacturers. However, certain aspects of design are published. Glass is a solid electrolyte, for which alkali-metal ions can carry current. The pH-sensitive glass membrane is generally spherical to simplify the manufacture of a uniform membrane. These membranes are up to 0.4 millimeters in thickness, thicker than original designs, so as to render the probes durable. The glass has silicate chemical functionality on its surface, which provides binding sites for alkali-metal ions and hydrogen ions from the solutions. This provides an ion-exchange capacity in the range of 10−6 to 10−8 mol/cm2. Selectivity for hydrogen ions (H+) arises from a balance of ionic charge, volume requirements versus other ions, and the coordination number of other ions. Electrode manufacturers have developed compositions that suitably balance these factors, most notably lithium glass. The silver chloride electrode is most commonly used as a reference electrode in pH meters, although some designs use the saturated calomel electrode. The silver chloride electrode is simple to manufacture and provides high reproducibility. The reference electrode usually consists of a platinum wire that has contact with a silver/silver chloride mixture, which is immersed in a potassium chloride solution. There is a ceramic plug, which serves as a contact to the test solution, providing low resistance while preventing mixing of the two solutions. With these electrode designs, the voltmeter is detecting potential differences of ±1400 millivolts. The electrodes are further designed to rapidly equilibrate with test solutions to facilitate ease of use. The equilibration times are typically less than one second, although equilibration times increase as the electrodes age.
=== Maintenance === Because of the sensitivity of the electrodes to contaminants, cleanliness of the probes is essential for accuracy and precision. Probes are generally kept moist when not in use with a medium appropriate for the particular probe, which is typically an aqueous solution available from probe manufacturers. Probe manufacturers provide instructions for cleaning and maintaining their probe designs. For illustration, one maker of laboratory-grade pH gives cleaning instructions for specific contaminants: general cleaning (15-minute soak in a solution of bleach and detergent), salt (hydrochloric acid solution followed by sodium hydroxide and water), grease (detergent or methanol), clogged reference junction (KCl solution), protein deposits (pepsin and HCl, 1% solution), and air bubbles.
=== Calibration and operation ===
The German Institute for Standardization publishes a standard for pH measurement using pH meters, DIN 19263. Very precise measurements necessitate that the pH meter is calibrated before each measurement. More typically calibration is performed once per day of operation. Calibration is needed because the glass electrode does not give reproducible electrostatic potentials over longer periods of time. Consistent with principles of good laboratory practice, calibration is performed with at least two standard buffer solutions that span the range of pH values to be measured. For general purposes, buffers at pH 4.00 and pH 10.00 are suitable. The pH meter has one calibration control to set the meter reading equal to the value of the first standard buffer and a second control to adjust the meter reading to the value of the second buffer. A third control allows the temperature to be set. Standard buffer sachets, available from a variety of suppliers, usually document the temperature dependence of the buffer control. More precise measurements sometimes require calibration at three different pH values. Some pH meters provide built-in temperature-coefficient correction, with temperature thermocouples in the electrode probes. The calibration process correlates the voltage produced by the probe (approximately 0.06 volts per pH unit) with the pH scale. Good laboratory practice dictates that, after each measurement, the probes are rinsed with distilled water or deionized water to remove any traces of the solution being measured, blotted with a scientific wipe to absorb any remaining water, which could dilute the sample and thus alter the reading, and then immersed in a storage solution suitable for the particular probe type.
== Types of pH meters ==
In general there are three major categories of pH meters. Benchtop pH meters are often used in laboratories and are used to measure samples which are brought to the pH meter for analysis. Portable, or field pH meters, are handheld pH meters that are used to take the pH of a sample in a field or production site. In-line or in situ pH meters, also called pH analyzers, are used to measure pH continuously in a process, and can stand-alone, or be connected to a higher level information system for process control. pH meters range from simple and inexpensive pen-like devices to complex and expensive laboratory instruments with computer interfaces and several inputs for indicator and temperature measurements to be entered to adjust for the variation in pH caused by temperature. The output can be digital or analog, and the devices can be battery-powered or rely on line power. Some versions use telemetry to connect the electrodes to the voltmeter display device. Specialty meters and probes are available for use in special applications, such as harsh environments and biological microenvironments. There are also holographic pH sensors, which allow pH measurement colorimetrically, making use of the variety of pH indicators that are available. Additionally, there are commercially available pH meters based on solid state electrodes, rather than conventional glass electrodes.
== History ==